beautypg.com

Myron L 760-A Aquaswitch User Manual

Page 15

background image

4.5.3 FULL SCALE CALIBRATION PROCEDURES
NOTE:

Before attempting to calibrate the 760-A Aquaswitch analog meter,

refer to Section 4.5.1.
STEP 1 Turn OFF main AC power.
STEP 2 Being careful no! to excessively strain the cables, unfasten

and remove the front panel.

STEP 3 Remove all five (5) wires from the 760-A Control board

terminal TB3. (See Fig. 4-1) .

STEP 4 By following the wiring diagram in Fig. 4-5, connect the

Calibrator Module’s lead wires to the 760-A Control board

TB3 terminal block connectors.

STEP 5 Turn ON the main AC power and verify that the 760-

A Aquaswitch analog meter is indicating a full scale

reading.

NOTE:

If the meter indicates an accurate reading, proceed to STEP 7. If

the meter does not indicate the appropriate reading, proceed to

STEP 6.
STEP 6 Turn the 760-A Control board’s Calibration trimmer (R8)

adjustment screw (see Fig. 4-1) until the meter displays

a full scale reading.

STEP 7 Remove the Calibrator Module’s lead wires and reconnect

the CS10 Cell cable wires as shown in Fig. 2-6.

4.6 PREVENTIVE CARE
The Myron L Company- recommends that the following Preventive

Care procedures be observed.

1 Try to prevent exposure to excessive heat and moisture.
2 The Aquaswitch main AC power source must be protected

against excessive voltage “spikes.”

3 Take care not to damage the Aquaswitch during handling.

NOTE:

Daily, weekly or monthly maintenance schedules are based upon

the frequency of use and the severity of the environment and

operating conditions.
STEP 1 Repeat the Aquaswitch Checkout procedures to

verify satisfactory operation and/or isolate possible

troubleshooting symptoms.

STEP 2 Check all cable connections to insure that they are free

of moisture and contamination.

STEP 3 Inspect and replace damaged component boards and

cable assemblies.

4.7 ALTERNATIVE VALVING CONFIGURATIONS AND

OPERATION FOR CONTINUOUS PROCESS WATER

NOTE:

The following methods will provide an uninterrupted water supply

to process if a fresh Dl bank is NOT in place when the other Dl

bank exhausts. Refer to Fig. 2 - and 2-5 on page 5.
OPTION 1 Requirements:

Continuous process. No interruptions allowed.

Connections:

“A” output (TB2-5) to “A” valve AND “B” valve.

“B” output not connected.

Bleed not connected.

Omit bleed valve.

Configure “A” valve Normally Closed & “B” valve

Normally Open.

OPTION 2 Requirements:

Continuous process. Holding tank between Dl beds

and process.

Purge ok but no prolonged shutdown.

Connections:

Bleed output (TB2-4) connected to bleed valve.

Bleed configured as Normally Closed.

Other connections same as Option 1.

OPTION 3 Requirements:

Continuous process. Purge required only at system

startup.

Connections:

Same as Option 2.

Add manual valve in line with Bleed Valve.

13

This manual is related to the following products: