beautypg.com

Myron L 760-A Aquaswitch User Manual

Page 13

background image

4.3 “PURGE CYCLE”. CALIBRATION PROCEDURES
NOTE:

The Purge Cycle trimmer adjustment screw (R2) (see Fig. 4-1)

determines the time span that the Bleed and Process solenoid

valves are energized during either a Bank A or B Purge Cycle.
STEP 1 Being careful not to excessively strain the cable(s),

unfasten the enclosure’s front panel.

STEP 2 Turn the Logic Module trimmer adjustment screw (R2) to

the desired Purge Cycle setting. (See Fig. 4-3)

NOTE:

Turning the adjustment screw fully clockwise equals a minimum

Purge Cycle of approximately one (1) minute. Turning the

adjustment screw fully counterclockwise equals a maximum Purge

Cycle of nine (9) to thirteen (13) minutes.

11

This manual is related to the following products: