beautypg.com

Myron L 760-A Aquaswitch User Manual

Page 12

background image

SECTION 4

Component Identification,

Calibration and Preventive Maintenance

4.1 GENERAL
Section 4.2 provides detailed illustrations to assist the user in

identifying the Aquaswitch/760-A Aquaswitch primary components

and cable connector designations.
Sections 4.3 and 4.4 provide the user with the Aquaswitch

Purge Cycle calibration procedures and Alarm Reset jumper

adjustment.
Section 4.5 provides the user with the 760-A Aquaswitch Calibration

procedures.
Section 4.6 provides the user with the Myron L Company’s

recommended Preventive Care procedures.

4.2 PRIMARY COMPONENT IDENTIFICATION
As identified in Section 3, the Aquaswitch/760-A Aquaswitch switch

and indicator components are mounted directly to the front panel.

Component boards and the 760-A Aquaswitch analog meter are

mounted within its enclosure or to the back of the front panel.
NOTE:

When performing Calibration procedures, the user must take

extreme care to contact only the trimmer adjustment screws located

on the Aquaswitch component boards. Failure to do so could result

in damage to equipment and/or property. (See Fig. 4-1)

10

This manual is related to the following products: