beautypg.com

Calibration, 3 calibration – Pulsafeeder Pulsar Shadow User Manual

Page 23

background image

4. Start the pump at the zero stroke length setting and slowly increase the setting to 100 to prime

the pump. If this does not work, it will be necessary to fill the suction line.

5. Filling of the suction line will necessitate the use of a foot valve or similar device at the end

of the suction line so that liquid can be maintained above the reservoir level. Remove the
suction valve assembly, fill the line, replace the valve, then remove the discharge valve
assembly and fill the reagent head as described in Step (3) above. The pump will now self-
prime when started up per step (4) above.


6.3

Calibration

Figure 14


All metering pumps must be calibrated in order to accurately specify stroke length settings for
required flow rates. For pumps provided with DLC/M electronic controls, refer to separate
instructions provided with those controllers.

A typical calibration chart is shown in Figure 14. Although output is linear with respect to stroke
length setting, an increase in discharge pressure decreases output uniformly, describing a series of
parallel lines, one for each pressure (only two are shown).

The theoretical output flow rate at atmospheric discharge pressure is based on the displacement of
the diaphragm, stroke length and the stroking rate of the pump. With increasing discharge
pressure there is a some corresponding decrease in output flow. Pumps are rated at a certain flow
at their rated pressure (check nameplate). Whenever possible, calibration should be performed
under actual process conditions (i.e., the same or a similar process liquid at system operating
pressure).

To construct a calibration chart, measure the flow rate several times at three or more stroke
settings (i.e., 25, 50, 75, and 100), plot these values on linear graph paper, and draw a best-fit line
through the points. For stable conditions, this line should predict settings to attain required
outputs.

17