beautypg.com

A&D Waterproof Digital Platform Scale HW-60KV-WP User Manual

Page 53

background image

HV-WP/HW-WP Series

Page 51

16. RS-232C Serial Interface

16.3.

16.3.

16.3.

16.3. Command mode

Command mode

Command mode

Command mode

The command mode is the function which can perform "output data", "controlling the
scale" and "setting parameters" by a command transmitted from a computer

Caution

Allow at least 500 milliseconds between commands.

16.3.1.

16.3.1.

16.3.1.

16.3.1. Command List

Command List

Command List

Command List

The following explanation uses "format 1 ( f15 0 )"

Data output

The current weighing data is output.
Template

Q

Command
Reply

Selection of mode and unit

Selects the mode and unit. This is the same as the

MODE

switch.

Template

U

Command
Reply

The scale changes mode and unit.

Zero

The current mass value is set to the zero point.
This is the same as the

ZERO

switch.

Template

Z

Command
Reply

The mass value becomes zero and the zero point mark is displayed.

Tare

The current mass value is set to zero after placing a tare (container, receptacle, case,
etc) and the net value is displayed. This is the same as the

TARE

switch.

Template

T

Command
Reply

The current mass value becomes zero and the net mark is displayed.

Cancel of tare

The display value becomes the gross and the net mark is turned off. (The tare value
becomes zero.)
Template

CT

Command
Reply

Gross is displayed and the net mark is turned off.