A&D Waterproof Digital Platform Scale HW-60KV-WP User Manual

Page 31

background image

HV-WP/HW-WP Series

Page 29

11. Upper/Lower Comparator Function

Function table

Meaning and purpose

A

f6 7

When the display value becomes stable and not nearly-zero, the
scale compares the value.

A
A

Caution

The upper limit value (HI) must be greater than the lower limit value (LO).
The parameters of the upper limit value (HI) and the final value (HI) use the
same memory. The parameters of the lower limit value (LO) and the zero band
(LO) use the same memory.
The upper/lower comparator function, the simple batch function and the
full/dribble batch function can not be used at the same time because these
parameters use common memory.

11.1.

11.1.

11.1.

11.1. Preparation (Setting Parameters)

Preparation (Setting Parameters)

Preparation (Setting Parameters)

Preparation (Setting Parameters)

Step 1 Turn off the display.

Press the

ON/OFF

switch while the

ZERO

switch is

pressed and held. The function table is displayed.

Step 2 Press the

ENTER

switch to display an item of the

accumulation function (

f6

).

Step 3 Select a parameter of the comparator function

(

f6 0

~

f6 7

) with the

switch.

Step 4 Press the

ENTER

switch to store the new parameter.

Step 5 Press the

F

switch and the

ENTER

switch to exit from

the function table. Then the scale displays the

weighing mode.