A&D Waterproof Digital Platform Scale HW-60KV-WP User Manual
Page 44
Page 42
HV-WP/HW-WP Series
14. Calibration (Adjusting Scale)
14.2.3.
14.2.3.
14.2.3.
14.2.3. Calibration of the Zero Point
Calibration of the Zero Point
Calibration of the Zero Point
Calibration of the Zero Point
Step 8 Press and hold the
CAL
switch to enter
the calibration mode after displaying
normal weighing for 30 minutes.
Then
Cal 0
is displayed
Step 9 With nothing on the weighing pan, press
the
ENTER
switch while the stable mark is
displayed. The scale stores the current
condition as the zero point.
Step 10 The scale displays
5pn1
for several
seconds. Finishing the calibration mode
at this stage, proceed to step 14.
14.2.4.
14.2.4.
14.2.4.
14.2.4. Span Calibration
Span Calibration
Span Calibration
Span Calibration
Step 11 Set the value of the calibration mass
using the following switches. (This initial
value is according to each product.)
∧
switch
Selecting the number of a
figure.
<
switch
Selecting a figure.
Step 12 Place the mass on the weighing pan
which was set at step 11, press the
ENTER
switch while the stable mark is
displayed. The scale then calculates the
span and stores it.
Step 13 The scale displays
end
at the finish.
Remove the mass from the weighing pan.
Step 14 Press the
CAL
switch to return to the normal
weighing mode.