beautypg.com

Step 6 – changing the calibration coefficients, Step 7 – addressing the probe – Campbell Scientific Sentek TriSCAN® Manual v 1.2a User Manual

Page 21

background image

TriSCAN Manual Version 1.2a

Copyright © 1991 – 2004 Sentek Pty Ltd All rights reserved

Page 14

3. Repeat this process for each sensor on the probe.


The expected air and water counts are dependent on the sensor type and sampling mode. Normalization
values outside the expected range may indicate errors in the normalization steps or problems with the
hardware.

Table 2. Expected air and water counts for different sensors

Sensor Type

Sampling Mode

Air Range

Water Range

EnviroSCAN sensor (white)

Moisture

36000±3000 23000±3000

TriSCAN sensor (blue)

Moisture

33000±2000 22800±2000

Salinity

20500±2000 13500±2000

EasyAG sensor (white)

Moisture

44000±2000 31000±2000

Moisture (sensor at bottom) 50000±2000 32000±2000

EasyAG with TriSCAN sensor (blue) Moisture

33000±2000 25000±2000

Salinity

20500±2000 15000±2000

Step 6 – Changing the Calibration Coefficients







To change the calibration coefficients of the sensors:

1. Click on the sensor coefficients cell when it is selected and it will go into edit mode as shown above.
2. Type the new coefficients separated by semicolons. To accept the new coefficients click outside the

cell or press Enter. If you want to discard the changes while in edit mode then press Escape. The
new coefficients will not be set in the probe’s configuration until you use the Write to Probe button.


The default calibration equation coefficients provided for different sensor types are listed in Table 3.

Table 3. Default Calibration Coefficients

Sensor Type

A

B

C

Moisture sensors 0.1957 0.404 0.02852

TriSCAN sensors

1.0

1.0

0

Step 7 – Addressing the Probe

Where there are multiple probes connected to a logger in a multi-drop situation, it is necessary to assign
each probe a unique address to avoid a clash in communication.