Page 25: Ranges

background image

OBS-3+ and OBS300 Suspended Solids and Turbidity Monitors

NTU value with a 2

nd

-order polynomial. This section explains how to do a

turbidity calibration.

0

1000

2000

3000

4000

Turbidity (TU)

0.0

0.2

0.4

0.6

0.8

1.0

N

or

m

al

iz

ed

O

B

S-

3+

R

es

po

ns

e

(O

PV

33

0

VC

SE

L)

FIGURE 8-1. Normalized response of OBS-3+ to AMCO Clear

®

turbidity.

The inset shows the response function of an OBS sensor

to high sediment concentrations.

GFS Chemicals manufactures AMCO Clear

®

SDVB turbidity standards that

are used to calibrate an OBS sensor. SDVB standards are made for individual

instruments and standards made for one model of turbidity meter cannot be

used to calibrate a different model. The standard values for calibrating OBSs

are: 125 NTU (GFS Part No. 8428), 250 NTU (GFS Part No. 8429), 500 NTU

(GFS Part No. 8430), 1000 NTU (GFS Part No. 8431), 2000 NTU (GFS Part

No. 8432), and 4000 NTU (GFS Part No. 8433). The standard values and

volumes required for the standard low ranges are given in TABLE 8-1. SDVB

standards have a shelf life of two years provided that they are stored in tightly

sealed containers and evaporation is minimized.

TABLE 8-1. SDVB NTU values for turbidity calibrations

in standard low ranges.

Low Range

1st Std. Value

2nd Std. Value

3rd Std. Value

250

125 (1000 ml)

250 (1000 ml)

1000 (500 ml)

500

250 (1000 ml)

500 (500 ml)

2000 (500 ml)

1000

500 (500 ml)

1000 (500 ml)

4000 (500 ml)

0

20000

40000

60000

SSC (mg/l)

0

1000

2000

3000

4000

Tu

rb

id

ity

(N

TU

)

A B

C

15