CVS4200 / BVS4300 Stationary Samplers

The controller has now been given a new
value for PINCH VALVE. The new value
resides in the NEW ENTRIES area of the
controller memory. To make these changes
active, press the RESTART key; press it
again to confirm your choice. The controller
will then be set to perform in conjunction
with the parameters programmed under the
START DELAY, SAMPLE INITIATION
and PROGRAM TYPE settings.


Summary of Sequence:
SET, OTHER OPTIONS, ARROW(S), ENTER, ARROW(S), ENTER,
RESTART, RESTART.

7.6.7.5 OTHER OPTIONS - Fault Shutdown

The following sequence of entries are made
on the Touchpad to describe how the
Sampler controller is to operate some of the
hardware specified in its configuration. The
ACTIVE SETTINGS are not being altered.

1. Press the SET key.

2. Press the OTHER OPTIONS key.

3. Press an ARROW key. Continue until
Fault Shutdown is shown on the display.

4. Press the ENTER key.


Fault Shutdown is used to control whether the sampler will cease taking
samples after a predetermined number of unsuccessful attempts.