The display will echo the last entry with
Any wrong entries will require re-entry. If
ENTER has not been pressed, reposition the
prompt over the incorrect digit and replace
it. After ENTER has been pressed, the entire
entry must be redone from the beginning
(press SET).
The controller has now been given a new
value for SAMPLE INITIATION. The new
values reside in the NEW ENTRIES area of
the controller memory. To make these
changes active, press the RESTART key;
press it again to confirm your choice. The
controller will then wait until the required
pulses have been received before taking a
sample.
Summary of Sequence:
SET, SAMPLE INITIATION, ARROW(S), ENTER, #######, ENTER,
RESTART, RESTART.
7.6.5.5 SAMPLE INITIATION using External Contact
The following sequence of entries are made
on the Touchpad to form a basic operating
parameter for operation of the sampler. This
will determine the time between samples
being taken. The ACTIVE SETTINGS are
not being altered.
1. Press the SET key.
2. Press the SAMPLE INITIATION key.
3. Press an ARROW key. Continue until
External Contact is shown on the display.
4. Press the ENTER key.