3 program type daily cycle, Program type - daily cycle – Campbell Scientific CVS4200 / BVS4300 Stationary Samplers User Manual
Page 62

CVS4200 / BVS4300 Stationary Samplers
7.6.6.3 PROGRAM TYPE - Daily Cycle
The following sequence of entries are made
on the Touchpad to describe how the
Sampler controller is to store the samples it
takes, in the hardware specified in its
configuration. The ACTIVE SETTINGS are
not being altered.
1. Press the SET key.
2. Press the PROGRAM TYPE key.
3. Press an ARROW key. Continue until
Daily Cycle is shown on the display (for
storage in a single container or a multiple
container).
4. Press the ENTER key.
The display will respond with the message
“DAILY CYCLE Total Bottles ‘nn’”. The
two digits will be flashing as a prompt; they
can be changed using the arrow keys.
Press ENTER. The second line of the
display will change to show a day of the
week. The first day displayed will be the
day the programming is being done.
Following the day of the week will be a
flashing digit, which is prompting for a
numeric input. Using the number keys (0-9)
enter the number of bottles to be utilized on
the displayed day, to a maximum of 9 or the
total number of containers not yet allocated
from the array.
Press ENTER. This will register the value
for the displayed day and advance the
display to the next day. Again, enter a
number for that particular day, remembering
that only containers not previously allocated
can be chosen. If no change in the displayed
54