6.4 Vacuum System Specifications
TABLE 6-5. Vacuum System Specifications
Feature Description
Switches
Run/Off (SPST Toggle).
Power On/Off (5 A lighted breaker).
Refrigerator On/Off (5 A lighted breaker).
Heater On/Off (5 A lighted breaker).
Sample Volume
Adjustable, 50 to 500 cc
Adjustable, 50 to 1,000 cc
Maximum Horizontal
Transport Distance
76.2 m (250 ft); assumes no vertical lift
Maximum Vertical
Lift
8.2 m (27 ft) for 3/8 inch system
6.1 m (20 ft) for 5/8 inch system
Metering Chamber
Cover
Nylon
Volume Control Tube
316 stainless steel
Metering Chamber
Level Electrode
316 stainless steel
Intake Hose Material
Nylon-Reinforced PVC
Discharge Hose
Material
Latex
Refrigerator
Small (composite): 4.4 cu ft, adjustable to 4ºC.
(optional)
Large (discrete): 5.8 cu ft, adjustable to 4ºC, glass-
door, digital display. (optional)
6.5 Sample Transport Velocity
TABLE 6-6. Vertical Velocity
Height
System Size
0 m
(0 ft)
1.5 m
(5 ft)
3.1 m
(10 ft)
4.6 m
(15 ft)
5.5 m
(18 ft)
6.1 m
(20 ft)
6.7 m
(22 ft)
7.6 m
(25 ft)
8.2 m
(27 ft)
3/8 inch
2.16 m s
-1
(7.1 ft s
-1
)
2.16 m s
-1
(7.1 ft s
-1
)
1.83 m s
-1
(6 ft s
-1
)
1.52 m s
-1
(5 ft s
-1
)
1.34 m s
-1
(4.4 ft s
-1
)
1.25 m s
-1
(4.1 ft s
-1
)
1.10 m s
-1
(3.6 ft s
-1
)
0.91 m s
-1
(3 ft s
-1
)
0.79 m s
-1
(2.6 ft s
-1
)
5/8 inch
(composite
samplers
only)
1.52 m s
-1
(5 ft s
-1
)
1.40 m s
-1
(4.6 ft s
-1
)
1.19 m s
-1
(3.9 ft s
-1
)
0.94 m s
-1
(3.1 ft s
-1
)
0.82 m s
-1
(2.7 ft s
-1
)
0.55 m s
-1
(1.8 ft s
-1
)
0 m s
-1
(0 ft s
-1
)