beautypg.com

6 programming program type, 1 program type overview, Programming program type – Campbell Scientific CVS4200 / BVS4300 Stationary Samplers User Manual

Page 59: Program type overview

background image

CVS4200 / BVS4300 Stationary Samplers

The display will echo the last entry with
.

The controller has now been given a new
value for SAMPLE INITIATION. The new
value resides in the NEW ENTRIES area of
the controller memory. To make this change
active, press the RESTART key; press it
again to confirm your choice. The controller
will then wait until a contact closure has
been received before taking a sample.


Summary of Sequence:
SET, SAMPLE INITIATION, ARROW(S), ENTER, RESTART,
RESTART.

7.6.6 Programming PROGRAM TYPE

7.6.6.1 PROGRAM TYPE Overview

PROGRAM TYPE is the function that determines how the sampler will
perform its program. A variety of options are available. These options are
programmable from the Touchpad and require only that the sampler have the
correct equipment configuration to utilize them. The basic function of the
PROGRAM TYPE is to determine the movement of the distributor.

The program option which determines that
all the samples that are gathered will be
placed in a single container. The sampler
program terminates after the specified
number of samples have been taken.

This option is used for discrete sampling
applications, to deposit one or a number of
samples to one container before advancing
to the next container. The interval betwee
samples is controlled by the SAMPLE
INITIATION
function. The
multi-composite setting is programmable up
to 99 samples per container, for up to 24
containers depending on the hardware
configuration.

n

51