3 sample initiation using pulse input, Sample initiation using pulse input – Campbell Scientific CVS4200 / BVS4300 Stationary Samplers User Manual
Page 55

CVS4200 / BVS4300 Stationary Samplers
When the five digits are entered, press
ENTER. Any wrong entries will require re-
entry. there are two methods of correcting a
mistake. The digit flashing “wraps around”
and begins again, at which time the correct
entry may be pressed. Alternately, the
arrow keys can be used to reposition the
flashing prompt over the error, which can
then be replaced with the correct value.
The display will echo the last entry with
The controller has now been given a new
value for SAMPLE INITIATION. The new
values reside in the NEW ENTRIES area of
the controller memory. To make these
changes active, press the RESTART key;
press it again to confirm your choice. The
controller will then wait the designated time
before taking a sample.
Summary of Sequence:
SET, SAMPLE INITIATION, ARROW(S), ENTER, ###:##, ENTER,
RESTART, RESTART.
7.6.5.3 SAMPLE INITIATION using Pulse Input
The following sequence of entries are made
on the Touchpad to form a basic operating
parameter for operation of the sampler. This
will determine the number of pulses from the
start of a sample cycle to the start of the next
sample cycle. The ACTIVE SETTINGS are
not being altered.
1. Press the SET key.
2. Press the SAMPLE INITIATION key.
47