beautypg.com

2 test for wind offset, Test for wind offset, Csat3 – Campbell Scientific CSAT3 3-D Sonic Anemometer User Manual

Page 50

background image

CSAT3 Three Dimensional Sonic Anemometer

The standard datalogger instruction reports either the speed of sound (c), option

90 or 99, or the sonic temperature (T

s

), Command/Option 91 or 98, cannot be

used with a CSAT3 running a cold shifted calibration. Instead, use option 92

“Trigger and Get Wind & c-340 Data” or option 97 “Get Wind and c-340 Data

After a Group Trigger” and then add 337 m s

–1

. To find T

s

, use Equation 9 in

Appendix C.

11.2.2 Test for Wind Offset

To test wind offset on a CSAT3, create an environment where there is no wind.

It is difficult to do this in the field, thus it is best to collect CSAT3 wind offset

data in a field office or the lab. A zero wind environment can be created with a

kitchen waste bin liner.

Mount the head in the normal horizontal fashion, or hang the head from the

block that supports the CSAT3 upper and lower arms. Do not lay the CAST3

head on its side or balance it on the transducers.

Cover the CSAT3 head with a medium (13 gallon) kitchen waste bin liner.

Fold, tape, or tie closed the opening of the liner to prevent air from moving in

and out of the liner. Ensure that the liner does not obstruct any of the three

CSAT3 sonic paths.

Connect the CSAT3 power cable to a 12 Vdc power supply. Connect the

CSAT3’s RS-232 port to the PC’s RS-232 port using the RS-232 cable that

shipped with the CSAT3. This cable was originally located underneath the

circular “cut out” in the CSAT3 case.

Start the CSAT3 PC support software (CSAT32.EXE) to establish

communications between the PC and the CSAT3. A copy of the software is

available from the Campbell Scientific web site, in the Support|Downloads

section. If successful communications are established, the “Status” window

will report “Good Data”. To prevent erroneous measurements due to ultrasonic

reflection within the plastic bag, select a 1 Hz Measurement Rate (FIGURE

11-3).

FIGURE 11-3. CSAT3 running at 1 Hz and successfully communicating

with the CSAT3

Double click on the vertical axis and set it up as shown in FIGURE 11-4.

Next, double click on the vertical axis label and set it up to report 2 significant

digits (FIGURE 11-5).

40