Campbell Scientific CSAT3 3-D Sonic Anemometer User Manual
Page 31

CSAT3 Three Dimensional Sonic Anemometer
Command/Option 90, 91, 98, or 99 must not be used to
collect data from a CSAT3 running a cold shifted
calibration, see Section 11.2.1.
Option 1, 2, 3, 5, 6, 10, 12, 15, 20, 30, and 60: “Set Execution Parameter”
tells the CSAT3 which measurement parameters to use and what frequency to
expect the measurement trigger from the datalogger. The datalogger Execution
Interval and the appropriate CSAT3 Execution Parameter is summarized in
Option 61 and 62: “Set Execution Parameter” tells the CSAT3 to oversample
(Section 8.1.2) the wind measurements at a rate of 60 Hz and expect a trigger
at a rate of 10 Hz (option 61) or 20 Hz (option 62) from a CR3000, CR1000,
CR5000, CR800/850, CR9000(X), CR23X or 21X datalogger.
Option 51 and 52: “Set Execution Parameter” tells the CSAT3 to oversample
(Section 8.1.2) the wind measurements at a rate of 48 Hz and expect a trigger
at a rate of 8 Hz (option 51) or 16 Hz (option 52) from a CR10(X) datalogger.
TABLE 10-3. CR3000, CR1000, CR5000, CR800/850, CR9000(X), CR23X and 21X Execution Interval
and CSAT3 Execution Parameter
Datalogger Execution Interval
(seconds)
Datalogger Execution Frequency
(Hz)
CSAT3 Execution Parameter
0.05
20
20
0.0625
16
15
0.075
13.333
12
0.0875 to 0.1
11.429 to 10
10
0.1125 to 0.1625
8.889 to 6.154
6
0.175 to 0.2
5.714 to 5
5
0.2125 to 0.325
4.706 to 3.077
3
0.3375 to 0.5
2.963 to 2
2
0.5125 to 1
1.951 to 1
1
TABLE 10-4. CR10(X) Execution Interval and CSAT3 Execution Parameter
Datalogger Execution Interval
(seconds)
Datalogger Execution Frequency (Hz)
CSAT3 Execution Parameter
0.0625
16
15
0.078125
12.8
12
0.09375
10.667
10
0.109375 to 0.15625
9.143 to 6.4
6
0.171875 to 0.1875
5.818 to 5.333
5
0.203125 to 0.328125
4.923 to 3.047
3
0.34375 to 0.5
2.909 to 2
2
0.515625 to 1
1.939 to 1
1
CAUTION
21