11.2 Calibration

The CSAT3 is calibrated over the temperature range of –30 to +50 °C

(standard) or –40 to +40 °C (cold shifted). The CSAT3 may or may not make

measurements outside the calibration range. Any measurements taken outside

the calibration range will be suspect. The CSAT3 does not require any field

calibration.

The CSAT3 electronics box contains unique calibration

information for its respective sensor head. Do not

mismatch CSAT3 sensor heads and electronics. Doing

so will result in erroneous wind and speed of sound

measurements.

If the CSAT3 requires a calibration, it will develop a wind offset greater than

the specifications, or it will set diagnostic flags, under dry conditions with little

to no wind and with no obstruction in the ultrasonic paths. CSAT3s running

embedded code version 4 will develop the offset first and then possibly set

diagnostic flags under dry conditions. CSAT3s running embedded code

version 3 will set diagnostic flags before a wind offset develops.

11.2.1 Cold Shifted Calibration

A CSAT3 with a cold shifted calibration is calibrated over the temperature of

–40 to +40 °C. To improve the speed of sound reporting resolution, the

CSAT3 sends the speed of sound minus an offset to a datalogger via the SDM

interface or a PC via the RS-232 interface. The data acquisition system must

add the appropriate offset to determine the speed of sound. For the standard

temperature range the offset 340 m s

–1

. This offset is added by the

CSAT3()/SDM-CSAT3 (P107) instruction Command/Option 90, 91, 98, and

99. For the cold shifted calibration range, the offset is 337 m s

–1

.

WARNING