1. Function

The 4WFBS120, 4WFBS350, and 4WFBS1K Terminal Input Modules (TIM)
complete a full Wheatstone bridge for a single strain gage or other sensor that
acts as a single variable resistor. The difference between the three models is in
the resistor that matches the nominal resistance of a 120 ohm, 350 ohm, or
1000 ohm quarter bridge strain gage. It can also be used to complete the back
half of a Wheatstone bridge for use in a ¼ bridge strain circuit (1 active
element) using a dummy gage, or in a ½ bridge strain circuit (2 active
elements).

FIGURE 1-1. Terminal Input Module with CR1000

2. Specifications

2:1 Resistive Divider
Resistors:

1 k

Ω/1 kΩ

Ratio tolerance @ 25

°C:

±0.01%

Ratio temperature coefficient:

0.5 ppm/

°C

(-55

°C to 85°C)

Power rating per element:

0.1 W @ 70

°C


Completion Resistor: 120, 350, or 1000

Ω

Tolerance

@

25

°C:

±0.01%

Temperature

coefficient:

±0.8 ppm

°C

-1

(-55

°C to 85°C)

Power

rating:

0.25 W @ 70

°C