Calibration
Settings
Sets/Reads
Calibration
related variables.
Argument 1 = T
Hours since last
time unit was
calibrated.
NOTE: has to be
reset to zero after
calibration.
19
CS
S
Meter Size/Pipe ID
[inches]
alue> [inches] CSS: Example: CSS:0.75 No Argument (Returns Current Meter Size settings) CSS: Example: CSS:0.75 F Meter Full Scale Range in liter/min alue> [liter/min] CSF: Example CSF:1000.0 No Argument (Returns Current Meter FS Range) CSF: Example: CSF:1000.0 C Meter Calibration Factor Hz/liter/min] (for STD condi- tions) alue> [Hz/l/min] CSC: Example CSC:1.0 No Argument (Returns Current Meter Calibration Factor) CSC: Example CSC:1.0 O Meter Low and High Frequency Cut Off [Hz] alue> (Hz) V alue>(Hz) CSO: alue>, : alue> No Argument (Returns Current Cut off V alues settings) CSO: alue>, : alue> L Meter Low Flow Cut Off in % of full scale alue> (%FS) CSL: Example: CSL:5.0 No Argument (Returns Current value) CSL: Example: CSL:5.0 T Calibration/Mainten ance T imer No Argument (read timer) CST : Example: CST :1024.2 Z Set Cal. T imer to Zero CST :Z