Aalborg Vortex Insertion User Manual
Page 101
96
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Argument 1 = R
Read DL record
in specified index.
Returns Date
/T
ime stamp and
process variable
value only if valid
data are present.
Process variable
value in current EU.
S
Read DL Status Info:
Current Index
[0- 15871]
Rollover counter
No Argument
DLS:
Example:
DLS:1057,0
R
Read DL record
alue> Record Index: [0-15871] DLR: ime>, value> Example:DLR: 07/14/2010,16:42:17,25.6 Z Set index to Zero DLZ Flowing Fluid Sets/Reads currently active flowing fluid. See below list of supported Liquids and Gases. 24 FF alue> Fluid type: 0 or 1 0 – Liquids 1 – Gases alue> Fluid Index: Liquids: [0-4] Gases: [0-12] FF: Example: FF:0,2 No Argument (Returns Current settings) FF:0,2,AIR Goyal-Dorais equation related parameters (other liquids) Sets/Reads Goyal-Dorais equation related parameters (for other liquids). 25 GD alue>Molecular W eight Mw (Lb/LbMol) [0-99999] alue> Crit. T emp T c (deg. F) [-459–99999] alue> Crit. Pressure Pc (PSIA) [0-99999] alue> Critical Compressibility factor Zc [0.0 – 1.0] GD: GD:18.1,705, 3197.9,0.23 No Argument (Returns Current settings) GD: GD:18.1,705, 3197.9,0.23 API2540 equation related parameters (Oil) Sets/Reads API2540e quation related parameters (for Oil). 26 AP alue> Density at 60F (kg/m^3) [0.0-99999.0] alue> K0 (no units) [0.0-9999.9] alue> K1 (no units) [0.0–99.9] AP: Example: AP:898.2,341.09,0.0 No Argument (Returns Current settings) AP: AP:898.2,341.09,0.0