Aalborg Vortex Insertion User Manual
Page 88
![background image](/manuals/414129/88/background.png)
83
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Te
mperature
Alarms**
Sets / reads the status of the
temperature alarms.
Note: High and Low limits have
to be entered in the °C. High
alarm value has to be more
than Low alarm value.
Alarm conditions:
Temp. > High Limit = H
Temp. < Low Limit = L
Low < T
emp. < High = N
9
TA
H
(high temp. limit)
alue> (-20.1-260 °C) TA H: alue [°C]> L (low temp. limit) alue> (-20.1-260 °C) TA L: alue [°C]> A (action delay in seconds) alue> (0-3600 sec.) TAA: alue [Sec]> E (enable alarm) TA :E D (disable alarm)* TA :D R (read current status) TA R:N (no alarm) TA R:H (high alarm) TA R: L (low alarm) S (Read current settings) TAS:M,L,H,D,B where: M – mode (E/D) L – Low settings (°C) H – High settings (°C) D – Action Delay (sec) B – Latch mode (0-1) B Block (Latch) mode alue> (0-disabled*) (1-enabl’d ) TAB: alue> where: V alue = 0 – 1