beautypg.com

Agilent Technologies 34401A User Manual

Page 66

background image

Autozero

(continued)

The following table shows the relationship between integration time and
autozero settings from the remote interface and the corresponding
front-panel settings.

From 34401A Specifications:

1

Remote Configuration

NPLC: 100
Autozero: On
Digits Displayed: 6

1

2

NPLC: 100
Autozero: Off
Digits Displayed: 6

1

2

NPLC: 10
Autozero: On
Digits Displayed: 6

1

2

NPLC: 10
Autozero: Off
Digits Displayed: 6

1

2

NPLC: 1
Autozero: On
Digits Displayed: 5

1

2

NPLC: 1
Autozero: Off
Digits Displayed: 5

1

2

NPLC: 0.2
Autozero: On
Digits Displayed: 5

1

2

NPLC: 0.2
Autozero: Off
Digits Displayed: 5

1

2

NPLC: 0.02
Autozero: On
Digits Displayed: 4

1

2

NPLC: 0.02
Autozero: Off
Digits Displayed: 4

1

2

Front-Panel Equivalent

Slow

6

digits

N/A

Fast 6 digits
Slow

5

digits

N/A

Slow

4

digits

N/A

N/A

Fast 5 digits

N/A

Fast 4 digits

Digits Displayed

N/A

6

1

2

N/A

6

1

2

N/A

5

1

2

N/A

5

1

2

N/A

4

1

2

Readings/Sec

N/A

0.6

N/A

6

N/A

60

N/A

300

N/A

1000

1

See the Agilent 34401A specifications listed on page 217.

Chapter 3 Features and Functions
Measurement Configuration

60