Execution Errors

-101

Invalid character
An invalid character was found in the command string. You may have
inserted a character such as #, $, or % in the command header or within
a parameter. Example:

CONF:VOLT#DC

-102

Syntax error
Invalid syntax was found in the command string. You may have
inserted a blank space before or after a colon in the command header,
or before a comma. Example:

SAMP:COUN ,1

-103

Invalid separator
An invalid separator was found in the command string. You may have
used a comma instead of a colon, semicolon, or blank space – or you may
have used a blank space instead of a comma. Example:

TRIG:COUN,1

or

CONF:FREQ 1000 0.1

-104

Data type error
The wrong parameter type was found in the command string. You may
have specified a number where a string was expected, or vice versa.
Example:

DISP:TEXT 5.0

-105

GET not allowed
A Group Execute Trigger (

GET

) is not allowed within a command string.

-108

Parameter not allowed
More parameters were received than expected for the command.
You may have entered an extra parameter, or you added a parameter
to a command that does not accept a parameter. Example:

READ? 10

-109

Missing parameter
Fewer parameters were received than expected for the command.
You omitted one or more parameters that are required for this
command. Example:

SAMP:COUN

5

Chapter 5 Error Messages
Execution Errors

173