Agilent Technologies 34401A User Manual
Page 181

-151
Invalid string data
An invalid character string was received. Check to see if you have
enclosed the character string in single or double quotes and that the
string contains valid
ASCII
characters. Example:
DISP:TEXT ’ON
(the ending quote is missing).
-158
String data not allowed
A character string was received but is not allowed for the command.
Check the list of parameters to verify that you have used a valid
parameter type. Example:
CALC:STAT ’ON’
-160 to -168
Block data errors
The multimeter does not accept block data.
-170 to -178
Expression errors
The multimeter does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (
GET
) or
*TRG
was received but the trigger
was ignored. Make sure the multimeter is in the “wait-for-trigger” state
before issuing a trigger, and make sure the correct trigger source is selected.
-213
Init ignored
An
INITiate
command was received but could not be executed because
a measurement was already in progress. Send a device clear to halt a
measurement in progress and place the multimeter in the “idle” state.
-214
Trigger deadlock
A trigger deadlock occurs when the trigger source is
BUS
and a
READ?
command is received.
5
Chapter 5 Error Messages
Execution Errors
175