[SENSe:]DETector:BANDwidth {3|20|200|MINimum|MAXimum}

Specify the lowest frequency expected in the input signal. The multimeter
selects the slow, medium (default), or fast ac filter based on the frequency
you specify.

MIN

= 3 Hz.

MAX

= 200 Hz. [Stored in volatile memory]

[SENSe:]DETector:BANDwidth? [MINimum|MAXimum]

Query the ac filter. Returns “+3.000000E+00”, “+2.000000E+01”, or
“+2.000000E+02”.

[SENSe:]ZERO:AUTO {OFF|ONCE|ON}

Disable or enable (default) the autozero mode. The

OFF

and

ONCE

parameters have a similar effect. Autozero

OFF

does not issue a new

zero measurement until the next time the multimeter goes to the
“wait-for-trigger” state. Autozero

ONCE

issues an immediate zero

measurement. [Stored in volatile memory]

[SENSe:]ZERO:AUTO?

Query the autozero mode. Returns “0” (

OFF

or

ONCE

) or “1” (

ON

).

INPut:IMPedance:AUTO {OFF|ON}

Disable or enable the automatic input resistance mode for dc voltage
measurements. With

AUTO OFF

(default), the input resistance is fixed

at 10 M

for all ranges. With

AUTO ON

, the input resistance is set to

>10 G

for the 100 mV, 1 V, and 10 V ranges. [Stored in volatile memory]

INPut:IMPedance:AUTO?

Query the input resistance mode. Returns “0” (

OFF

) or “1” (

ON

).

ROUTe:TERMinals?

Query the multimeter to determine if the front or rear input terminals
are selected. Returns “

FRON

” or “

REAR

”.

4