Agilent Technologies 34401A User Manual
Page 137

TRIGger:DELay {<
seconds
>|MINimum|MAXimum}
Insert a trigger delay between the trigger signal and each sample
that follows. If you do not specify a trigger delay, the multimeter
automatically selects a delay for you. Select from 0 to 3600 seconds.
MIN
= 0 seconds.
MAX
= 3600 seconds. [Stored in volatile memory]
TRIGger:DELay? [MINimum|MAXimum]
Query the trigger delay.
TRIGger:DELay:AUTO {OFF|ON}
Disable or enable an automatic trigger delay. The delay is determined
by function, range, integration time, and ac filter setting. Selecting a
specific trigger delay value automatically turns off the automatic trigger
delay. [Stored in volatile memory]
TRIGger:DELay:AUTO?
Query the automatic trigger delay setting. Returns “0” (
OFF
) or “1” (
ON
).
SAMPle:COUNt {<
value
>|MINimum|MAXimum}
Set the number of readings (samples) the multimeter takes per trigger.
Select from 1 to 50,000 readings per trigger.
MIN
= 1.
MAX
= 50,000.
[Stored in volatile memory]
SAMPle:COUNt? [MINimum|MAXimum]
Query the sample count.
TRIGger:COUNt {<
value
>|MINimum|MAXimum|INFinite}
Set the number of triggers the multimeter will accept before returning
to the “idle” state. Select from 1 to 50,000 triggers. The
INFinite
parameter instructs the multimeter to continuously accept triggers
(you must send a device clear to return to the “idle” state). Trigger count
is ignored while in local operation.
MIN
= 1.
MAX
= 50,000.
[Stored in volatile memory]
TRIGger:COUNt? [MINimum|MAXimum]
Query the trigger count. If you specify an infinite trigger count,
the query command returns “9.90000000E+37”.
4
Chapter 4 Remote Interface Reference
Triggering Commands
131