Measurement configuration commands – Agilent Technologies 34401A User Manual
Page 127

Measurement Configuration Commands
See also “Measurement Configuration,” starting on page 51 in chapter 3.
FUNCtion "<
function
>"
Select a measurement function. The function must be enclosed in quotes
in the command string (
FUNC "VOLT:DC"
). Specify one of the following
strings.
FUNCtion?
Query the measurement function and return a quoted string.
<
function
>:RANGe {<
range
>|MINimum|MAXimum}
Select the range for the selected function. For frequency and period
measurements, ranging applies to the signal’s input voltage, not its
frequency (use
FREQuency:VOLTage
or
PERiod:VOLTage
).
MIN
selects
the lowest range for the selected function.
MAX
selects the highest
range. [Stored in volatile memory]
<
function
>:RANGe? [MINimum|MAXimum]
Query the range for the selected function.
<
function
>:RANGe:AUTO {OFF|ON}
Disable or enable autoranging for the selected function. For frequency
and period, use
FREQuency:VOLTage
or
PERiod:VOLTage
.
Autorange thresholds: Down range at <10% of range; Up range
at >120% of range. [Stored in volatile memory]
<
function
>:RANGe:AUTO?
Query the autorange setting. Returns “0” (
OFF
) or “1” (
ON
).
VOLTage:DC
VOLTage:DC:RATio
VOLTage:AC
CURRent:DC
CURRent:AC
RESistance (
2-wire ohms
)
FRESistance (
4-wire ohms
)
FREQuency
PERiod
CONTinuity
DIODe
4
Chapter 4 Remote Interface Reference
Measurement Configuration Commands
121