The standard event register is cleared when:
•
You send a
*CLS
(clear status) command.
•
You query the event register using the
*ESR?
(event status register)
command.
The standard event enable register is cleared when:
•
You turn on the power and you have previously configured the
multimeter using the
*PSC 1
command.
•
You execute a
*ESE 0
command.
The standard event enable register will not be cleared at power-on if you
have previously configured the multimeter using
*PSC 0.
4
Chapter 4 Remote Interface Reference
The SCPI Status Model
141