background image

The standard event register is cleared when:

You send a

*CLS

(clear status) command.

You query the event register using the

*ESR?

(event status register)

command.

The standard event enable register is cleared when:

You turn on the power and you have previously configured the
multimeter using the

*PSC 1

command.

You execute a

*ESE 0

command.

The standard event enable register will not be cleared at power-on if you
have previously configured the multimeter using

*PSC 0.

4

Chapter 4 Remote Interface Reference
The SCPI Status Model

141