Sensor error detection limits

error indications

Sensor error detection (SE.BR, SE.SH):

Input

Range

Readout

Limit

CURR

Loop break (4..20 mA)

SE.BR

<= 3.6 mA; > = 21 mA

POTM

All, SE.BR on all 3­wire

SE.BR

> ca. 126 k

Ω

LIN.R

0...800

Ω

SE.BR

> ca. 875

Ω

0...10 k

Ω

SE.BR

> ca. 11 k

Ω

TEMP

TC

SE.BR

> ca. 750 k

Ω

/ (1.25 V)

RTD, 2­, 3­, and 4­wire

No SE.SH for Pt10, Pt20 and Pt50

SE.BR

> ca. 15 k

Ω

SE.SH

< ca. 15

Ω

Display readout below min.­ / above max. (­1999, 9999):

Input

Range

Readout

Limit

All

All

­1999

Display readout <­1999

9999

Display readout >9999

Readout at hardware error

Error search

Readout

Error cause

Test of internal CJC sensor

CJ.ER

CJC sensor defect or tem­

perature outside range

Checksum test of the setup in FLASH

FL.ER

Error in FLASH

Check measurement of analogue output current

AO.ER

1) Incorrect current output

(only S4...20/S20...4 mA)

Communications test PGMMOD00 / device

NO.CO

Connection error

Check that input signal matches input setup

IN.ER

1) Error levels on input

Check that saved setup in PGMMOD00 matches device

TY.ER

Setup mismatch

! Error indications in the display flash once per second. The help text explains the error.