This chapter provides the technical data of the DPT145.

Specifications

Table 12

Measured Parameters

Parameter

Measured Range

Dewpoint

-50 ... +30 °C (-58 ... +86 °F)

Pressure, absolute

1 ... 12 bar (14.5 ... 174 psi)

Temperature

-40 ... +80 °C (-40 ... +176 °F)

Table 13

Calculated Parameters

Parameter

Measured Range

Pressure, normalized to 20 °C (68 °F)

1 ... 12 bara (14.5…174 psia)

SF

6

or SF

6

/N

2

mixture density

0 ... 100 kg/m

3

ppm moisture, by volume

40 ... 40 000 ppm

Dewpoint, converted to atmospheric

pressure

-65 … +30 °C (-85 ... +86 °F)

Table 14

Performance

Property

Description / Value

Dewpoint accuracy

±3 °C (±5.4 °F), see Figure 12 on

page 55

Dewpoint stability

typical drift < 2 °C (3.6 °F)/5a

Pressure accuracy at 23 °C (73.4 °F)

±0.04 %FS

Pressure temperature dependence

±0.01 bar/10 °C (18 °F)

Temperature accuracy

0 ... +40 °C (+32 ... +104 °F)

±0.5 °C (± 0.9 °F)

-40 ... +80 °C (-40 ... +176 °F)

±1 °C (± 1.8 °F)

Density accuracy (pure SF

6

, 1 ... 10 bara)

0 ... 40 °C (+32 ... +104 °F)

±1 %FS

-40 ... +60 °C (-40 ... +140 °F)

±2.2 %

ppm accuracy, typical

(5 ... 1000 ppm, 7 bar)

±(7 ppm + 15 % of reading)

Sensor response time

Pressure response time

< 1 s

Dewpoint response time*

63% [90%] at 20°C and 1 bar

-50 -> -10 °C Tdf

5 s [10 s]

-10 -> -50 °C Tdf

10 s [2.5 min]

* system equilibrium related response time is typically longer