beautypg.com

Unknown serial settings, Technical support, On unknown serial settings – Vaisala DPT145 User Manual

Page 52

background image

User's Guide _______________________________________________________________________

52 ___________________________________________________________________ M211371EN-D

Unknown Serial Settings

If you are not sure what the current serial settings of your transmitter are,

follow the steps below to connect to the transmitter. This procedure

assumes you have the USB service cable available.

For more information on the USB service cable and the PuTTY terminal

application, see section Serial Communication on page 31.

1.

Connect the USB service cable to your PC, but do not connect it to

the transmitter yet.

2.

Start a terminal application and open a terminal session using the

default serial settings of 19200 8 N 1.

3.

Plug the USB service cable into the transmitter, and immediately

send the hash character # to the serial line:

#

4.

The normal startup of the transmitter is now interrupted, and the

transmitter is available for commands with the default serial

settings.

5.

You can now check the transmitter’s information by issuing

the ? command:

?

The transmitter’s currently stored serial settings are visible on the

line marked Baud P D S.

Technical Support

For technical questions, contact the Vaisala technical support by e-mail at

[email protected]

. Provide at least the following supporting

information:

- Name and model of the product in question
- Serial number of the product
- Name and location of the installation site
- Name and contact information of a technically competent person who