beautypg.com

Show firmware version, Show serial number, Serial line output commands – Vaisala DPT145 User Manual

Page 36: Start measurement output, Stop measurement output

background image

User's Guide _______________________________________________________________________

36 ___________________________________________________________________ M211371EN-D

Show Firmware Version

VERS

Example:

vers

DPT145 1.0.23

Show Serial Number

SNUM

Example:

snum

Serial number : F4040007

Serial Line Output Commands

Start Measurement Output

Use the R command to start the continuous outputting of measurement

values as an ASCII text string to the serial line. The format of the

measurement message is set with the FORM command.

R

Example (measurement message in default format):

r

Tdf= 12.5 'C Tdfatm= 13.6 'C H2O= 15561 ppm P= 0.947 bara

Pnorm= 0.943 bara Rhoo= 1.1 kg/m3 T= 24.5 'C

Tdf= 12.5 'C Tdfatm= 13.6 'C H2O= 15563 ppm P= 0.947 bara

Pnorm= 0.942 bara Rhoo= 1.1 kg/m3 T= 24.5 'C

Tdf= 12.5 'C Tdfatm= 13.5 'C H2O= 15555 ppm P= 0.948 bara

Pnorm= 0.943 bara Rhoo= 1.1 kg/m3 T= 24.5 'C

...

Outputting the results continues in intervals issued with the command

INTV. You can stop the output by pressing the Esc key, or with the

command S. Since the interface is half-duplex, you must enter the

commands when the transmitter is not outputting.

Stop Measurement Output

You can stop the measurement output by pressing ESC, or with the S

command:

S