Chapter 5 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 37

Set Output Interval

Use the INTV command to change the output interval of the

automatically repeating measurement messages. The measurement

messages are repeated in the RUN mode, or after the R command has

been given.

INTV [n xxx]

where
n

= time interval, range 0 ... 255

xxx

= time unit = "S", "MIN", or "H"

The shortest output interval (with n = 0) outputs the measurement

messages as quickly as the transmitter’s internal measurement cycle

produces them, without additional delay.

Example:

intv 1 min

Output interval: 1 MIN

Output a Reading Once

Use the SEND command to output a single measurement message. The

output message is sent in the format that has been set using the FORM

command.

SEND [address]

where
address = Address of the transmitter, range 0 ... 255.

Specifying the address is only necessary if the target

transmitter is in POLL mode, and has not been accessed

with the OPEN command.

Example:

send

Tdf= 15.4 'C Tdfatm= 16.2 'C H2O= 18539 ppm P= 0.959 bara

Pnorm= 0.951 bara Rhoo= 1.1 kg/m3 T= 25.4 'C