beautypg.com

Figure 10, Comparing dewpoint readings on mi70 – Vaisala DPT145 User Manual

Page 48

background image

User's Guide _______________________________________________________________________

48 ___________________________________________________________________ M211371EN-D

NOTE

In the procedure below, the readings of the transmitter and the reference

probe are checked simultaneously. You can also check the reference

probe and the DPT145 separately if they cannot be connected at the

same time.

Perform the check as follows:

1.

Place the reference probe and the DPT145 in the reference

environment.

2.

Connect the reference probe and the DPT145 to the MI70 indicator:

- Reference probe to Port I
- DPT145 to Port II

3.

Turn on the MI70 indicator.

4.

Check and adjust the environment settings if prompted. Refer to the

DM70 User’s Guide for the settings of the DMP74C probe.

5.

Disable the Auto power off function of the MI70 indicator to

prevent the indicator from powering off in the middle of the check.

Check the setting in the Settings - User Interface menu, and

change if necessary.

6.

The MI70 can display three parameters at one time. To compare

the readings of the DPT145 and the reference probe, select the desired

parameter from the DisplayQuantities and Units menu of the MI70.

Select the same parameter from port I and II, and the delta parameter that

shows the difference between the two. For example, ∆T

d/f

shows the

difference in dewpoint between the two probes.

1108-039

Figure 11

Comparing Dewpoint Readings on MI70

7.

After the measurement has stabilized, check the difference between

the readings. It will give you an indication of the approximate drift

of the DPT145 and its need for calibration. The difference between

the readings should be no greater than the combined measurement

uncertainty of the DPT145 and the reference probe.

8.

Write down the results and repeat the comparison for each

parameter you want to check.