Chapter 6 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 47

CHAPTER 6

MAINTENANCE

This chapter provides information that is needed in basic maintenance of

the DPT145.

Periodic Maintenance

Cleaning

The body of the transmitter can be cleaned by wiping with a moistened

lint-free cloth. Do not immerse the transmitter in liquid, and do not use

cleaning agents or solvents.

Calibration and Adjustment

DPT145 is fully calibrated as shipped from factory. If there is a reason to

believe that the device is not within the accuracy specifications, you can

perform a field check using the DM70 hand-held dewpoint meter, or an

MI70 indicator with another DPT145.

If the field check indicates that the DPT145 is not within its accuracy

specifications, contact a Vaisala Service Center or your local Vaisala

representative to have the DPT145 adjusted.

Field Check Using an MI70 Indicator

Vaisala recommends the following equipment for field checking:

- A fully charged MI70 indicator (the measurement display that is

included in the DM70 package)

- A calibrated reference probe that is suitable for the intended

measurement environment and dewpoint range. The DMP74C probe

is recommended for checking the dewpoint and temperature

measurement accuracy. For checking all parameters, you can use

another DPT145 transmitter as a reference.

- You also need a connection cable (Vaisala order code: 219980) to

connect the DPT145 to the MI70 indicator. If you have two DPT145

transmitters, you need two cables.