User's Guide _______________________________________________________________________

64 ___________________________________________________________________ M210865EN-F

Operating Environment

Temperature

-40 … +70 °C (-40 … +158 °F)

Relative humidity

0 … 100 %RH (up to +20 °C / +68 °F)

Pressure

0 … 50 bar (725 psia)

Measured gases

non-corrosive gases

Sample flow rate

no effect on measurement accuracy

Outputs

Two analog outputs (scalable)

0 … 20 mA

4 … 20 mA

0 … 5 V

0 … 10 V

Accuracy

voltage output

± 0.01 V

current output

± 0.01 mA

Typical temperature dependence

voltage output

± 0.001 % of span / °C

current output

± 0.005 % of span / °C

Digital output

RS485 (2-wire half-duplex)

ON/OFF output available instead of the 2

nd

analog output

purge inactive

0 V or 4 mA

purge active

5 V, 10 V or 20 mA

OR exceeded T

d

or ppm limit

user selectable

General

Sensor

Vaisala DRYCAP

®

180U

Thin-film capacitive polymer sensor

Startup time

10 .. 15 minutes

Recommended calibration interval

2 years

Operating voltage with

RS485 output

11* … 28 VDC

voltage output

15* … 28 VDC

current output

21 … 28 VDC

* For extended temperature down to -40 °C (-40 °F) or pressure up to
50 bar (725 psia) the supply voltage must be 21 ... 28 VDC.

Operating voltage fluctuation

max. 0.3 V

Supply current

normal measurement

20 mA + load current

during self-diagnostics

max. 220 mA pulsed

Load for

voltage output

min. 10 kOhm

current output

max. 500 Ohm