Chapter 5 ________________________________________________________ Serial Line Operation

VAISALA ________________________________________________________________________ 49

Serial Line Output Commands

Start Measurement Output

Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the FORM command.

R

Example (measurement message in default format):

r
Tdf= -2.5 'C H2O= 4919 ppm
Tdf= -2.5 'C H2O= 4918 ppm
Tdf= -2.5 'C H2O= 4917 ppm

Outputting the results continues in intervals issued with the command
INTV. You can stop the output by pressing the Esc key, or with the
command S. Since the interface is half-duplex, you must enter the
commands when the transmitter is not outputting.

Stop Measurement Output

You can stop the measurement output by pressing Esc, or with the S
command:

S

Set Output Interval

INTV [n xxx]

where

n

=

time interval = 0 ... 255

xxx

=

time unit = "S", "MIN", or "H"


Example:

intv 1 min
Output interval: 1 MIN

The shortest output interval (with n = 0) is approximately one second due
to the internal measurement cycle.