Calibrate analog output, Extend analog output range – Vaisala DMT152 User Manual

Page 54

background image

User's Guide _______________________________________________________________________

52 ___________________________________________________________________ M210865EN-F

Calibrate Analog Output

Use the ACAL command to force the analog outputs to the following
values:

- Current output: 2 mA and 18 mA

- Voltage output: 10 % and 90 % of the range

After entering the ACAL command, use a calibrated current/voltage
meter to measure the output, and enter the values. Note that the
calibration is only valid for the currently selected analog output modes.

ACAL [channel]

where

Channel

=

Analog output channel: 1 or 2

Example:

acal 1
Ch1:
I1 ( mA ) ? 0.166
I2 ( mA ) ? 18.1
3.61058569E+02 1.40927863E+03

Extend Analog Output Range

Use the AOVER command to allow the analog output channels to
exceed their specified range by 10%. The scaling of the quantity remains
as before; the extra range is used for additional measurement range in the
wet end.

AOVER [ON/OFF]

The following example illustrates how the analog output is affected.
Channel 1 outputs T

d

with voltage output 0 … 5 V (-80 °C … -30 °C).

After giving the AOVER ON command, the range is 0 … 5.5 V (-80 °C
… -25 °C). Note that the -30 °C T

d

point is still at 5 V.

Example:

aover on

: ON