Accuracy of calculated variables, Accuracy of dewpoint temperature °c, Accuracy of mixing ratio g/kg – Vaisala HMT360 User Manual
Page 85

Chapter 9 ____________________________________________________________ Technical Data
VAISALA_______________________________________________________________________ 83
Accuracy of Calculated Variables
Accuracy of the calculated variables depend on the calibration accuracy of
the humidity and temperature sensors; here the accuracy are given for ±2
%RH and ±0.2 °C.
Accuracy of Dewpoint Temperature °C
Accuracy of Mixing Ratio g/kg
(Ambient pressure 1013 mbar)
Relative humidity
Temp. 10
20
30
40
50
60
70
80
90
100
-40
1.86
1.03
0.76
0.63
0.55
0.5
0.46
0.43
—
—
-20
2.18
1.19
0.88
0.72
0.62
0.56
0.51
0.48
—
—
0
2.51
1.37
1
0.81
0.7
0.63
0.57
0.53
0.5
0.48
20
2.87
1.56
1.13
0.92
0.79
0.7
0.64
0.59
0.55
0.53
40
3.24
1.76
1.27
1.03
0.88
0.78
0.71
0.65
0.61
0.58
60
3.6
1.96
1.42
1.14
0.97
0.86
0.78
0.72
0.67
0.64
80
4.01
2.18
1.58
1.27
1.08
0.95
0.86
0.79
0.74
0.7
100
4.42
2.41
1.74
1.4
1.19
1.05
0.95
0.87
0.81
0.76
120
4.86
2.66
1.92
1.54
1.31
1.16
1.04
0.96
0.89
0.84
140
5.31
2.91
2.1
1.69
1.44
1.26
1.14
1.05
0.97
0.91
160
5.8
3.18
2.3
1.85
1.57
1.38
1.24
1.14
1.06
0.99
Relative humidity
Temp. 10
20
30
40
50
60
70
80
90
100
-40
0.003
0.003
0.003
0.003
0.003
0.004
0.004
0.004
—
—
-20
0.017
0.018
0.019
0.021
0.022
0.023
0.025
0.026
—
—
0
0.08
0.09
0.09
0.1
0.1
0.11
0.11
0.12
0.13
0.13
20
0.31
0.33
0.35
0.37
0.39
0.41
0.43
0.45
0.47
0.49
40
0.97
1.03
1.1
1.17
1.24
1.31
1.38
1.46
1.54
1.62
60
2.68
2.91
3.16
3.43
3.72
4.04
4.38
4.75
5.15
5.58
80
6.73
7.73
8.92
10.34
12.05
14.14
16.71
19.92
24.01
29.29
100
16.26
21.34
28.89
40.75
60.86
98.85
183.66 438.56 —
—
120
40.83
74.66
172.36 —
—
—
—
—
—
—