Analog output test, Technical support, Product returns – Vaisala HMT360 User Manual
Page 76: Technical support product returns, Table 7, Error codes, See section, For c, Reports an error; see

User’s Guide ______________________________________________________________________
74 ___________________________________________________________________ M010056EN-J
Analog Output Test
Analog outputs can be tested with dip switch 1, see
for details.
Technical Support
For technical questions, contact the Vaisala technical support by e-mail at
. Provide at least the following supporting
information:
-
Name and model of the product in question
-
Serial number of the product
-
Name and location of the installation site
-
Name and contact information of a technically competent person who
can provide further information on the problem.
Product Returns
If the product must be returned for service, see
www.vaisala.com/returns
.
For contact information of Vaisala Service Centers, see
www.vaisala.com/servicecenters
.
Table 7
Error Codes
Error Code
on Display
Error Code
on Serial Line
Typical Cause
RH error
RH error
HUMICAP® humidity sensor open
circut or short circut. Can be
caused by condensed water, or
damage to sensor or probe.
T error
T error
Pt100 temperature sensor open
circut or short circuit. Can be
caused by condensed water, or
damage to sensor or probe.
ADC error
Prb error
Probe module disconnected.
CPU EEPROM error EE error
Hardware failure in transmitter.
PRB EEPROM error EEP error
Hardware failure in probe module.