Installation in hazardous locations, Us and canadian requirements, European requirements – Vaisala HMT360 User Manual
Page 38: Category 1 (zone 0), Category 2 or 3 (zone 1 or 2), Us and canadian requirements european requirements, Category 1 (zone 0) category 2 or 3 (zone 1 or 2)

User’s Guide ______________________________________________________________________
36 ___________________________________________________________________ M010056EN-J
Installation in Hazardous Locations
US and Canadian Requirements
USA (FM): Wiring for intrinsically safe operation is shown in
B, Wiring for Intrinsically Safe Operation, FM, on page 91
Canada (CSA): Wiring for intrinsically safe operation is shown in
Appendix C, Wiring for Intrinsically Safe Operation, CSA, on page 93
.
European Requirements
CATEGORY 1 (Zone 0)
HMT360 has to be connected to Exia-certified associated apparatus with
galvanic isolation, gas group IIB or IIC.
CATEGORY 2 or 3 (Zone 1 or 2)
HMT360 has to be connected either to a Zener barrier or galvanic isolator
and
present examples of
galvanic isolators and Zener barrier connections (only Ch 1 connected).
NOTE
If both analog outputs are in use, the Ch 1 (-) and Ch 2 (-) must be short
circuited (see
NOTE
If both analog outputs are in use with a galvanic isolator, the Ch 1 (-) and
Ch 2 (-) must be short circuited (see