Water activity in jet fuel applications, Calculated variables (typical ranges), With hmp361 probe – Vaisala HMT360 User Manual
Page 79: Table 10, Table 11, Hmp361 calculated variables specifications, Table 12
Chapter 9 ____________________________________________________________ Technical Data
VAISALA_______________________________________________________________________ 77
Water Activity in Jet Fuel Applications
Calculated Variables (Typical Ranges)
With HMP361 Probe
With HMP363, HMP364, HMP365, HMP367 and
HMP368 Probes
Table 10
Water Activity in Jet Fuel Applications Specifications
Property
Description / Value
Measurement range
0 ... 1 a
w
at -40 ... +180 °C (-
40 ... +356 °F)
Accuracy when calibrated against high
quality, certified humidity standards
salt solutions (ASTM E104-85)
± 0.01 (0 ... 0.9)
± 0.02 (0.9 ... 1.0)
± 0.02 (0 ... 0.9)
± 0.03 (0.9 ... 1.0)
Response time (90 %) at +20 °C in still oil
(stainless steel filter)
10 min
Humidity sensor
Vaisala HUMICAP
®
Calculated variables available
mass concentration of water
relative humidity of saturation
saturation temperature
Accuracy of the water content
better than ± 15 % of the
reading
Table 11
HMP361 Calculated Variables Specifications
Property
Description / Value
Dewpoint temperature
-40 ... +60 °C
Mixing ratio
0 ... 160 g/kg d.a.
Absolute humidity
0 ... 160 g/m³
Wet bulb temperature
0 ... +60 °C
Table 12
HMP363, HMP364, HMP365, HMP367& HMP368
Probes Calculated Variables Specifications
Property
Description / Value
Dewpoint temperature
-40 ... +100 °C
Mixing ratio
0 ... 500 g/kg d.a.
Absolute humidity
0 ... 600 g/m³
Wet bulb temperature
0 ... +100 °C