Vaisala HMP240 User Manual
Page 87

HMP240
M210300en
Appendix 1: Serial commands
85
SMODE xxxx
aa
=
address of the transmitter
xxxx
=
STOP, RUN or POLL
The OPEN command sets the bus temporarily in STOP MODE so that the
SMODE command can be given. For example:
>SMODE
which mode is in use at the moment
Serial mode : STOP
>SMODE STOP
setting STOP mode
Serial mode : STOP
OPEN & CLOSE
OPEN nn
nn
=
address of the transmitter (0...99)
CLOSE
In STOP mode: command OPEN has no effect, CLOSE sets the transmitter in
POLL mode
In POLL mode: command OPEN sets the transmitter temporarily in STOP
mode, command CLOSE returns the instrument to POLL
mode
When more than one transmitter is connected to the same serial bus, the POLL
mode makes it possible to communicate with the transmitters. For example, a
relative humidity calibration is performed at transmitter 2 (
>OPEN 2
>CRH
...
>CLOSE