Vaisala HMP240 User Manual
Page 84

HMP240
Appendix 1: Serial commands
M210300en
82
Activating the date output
>fdate on
Form. date : ON
>r
1995-03-10 RH= 21.1 %RH T= 26.0 'C
1995-03-10 RH= 21.3 %RH T= 26.0 'C
1995-03-10 RH= 23.1 %RH T= 26.0 'C
1995-03-10 RH= 22.2 %RH T= 26.0 'C
1995-03-10 RH= 20.6 %RH T= 26.0 'C
...
Inactivating the date output
>fdate off
Form. date : OFF
>r
RH= 20.2 %RH T= 26.0 'C
RH= 19.9 %RH T= 26.0 'C
RH= 19.8 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
...
SERI
Serial bus settings
SERI b p d s x
b
=
bauds (300, 600, 1200, 2400, 4800, 9600)
p
=
parity (n = none, e = even, o = odd)
d
=
data bits (7 or 8)
s
=
stop bits (1 or 2)
x
=
duplex (H = half, F = full)
Giving the command on its own outputs the current settings:
>SERI
4800 E 7 1 FDX
The settings can be changed one parameter at a time or all parameters at once:
>SERI O H
changing parity and duplex
4800 O 7 1 HDX
>SERI 600 N 8 1 F
changing all parameters
600 N 8 1 FDX
NOTE
The serial bus settings become effective only after reset.