Vaisala HMP240 User Manual
Page 48

HMP240
User's Guide
M210300en
46
•
Insert the sensor head into the measurement hole of the
LiCl chamber.
•
Wait at least 10 minutes for the reading to stabilize.
•
Give command CRH
press
>CRH
RH : xx.x Ref1 ? yy.y
Press any key when ready...
•
If you want to see how the sensor stabilizes to the humid-
ity in the calibrator, enter c
RH : 11.9 Ref1 ? c
RH : 11.5 Ref1 ? c
RH : 11.5 Ref1 ? 11.3
Press any key when ready...
•
Insert the sensor head into the measurement hole of the
NaCl chamber.
•
Wait at least 20 minutes for the reading to stabilize.
•
Press any key, enter the second point value and press
RH : xx.x Ref2 ? yy.y
•
The stabilization of the sensor can be monitored by
entering c
6.1.2.2
With display / keypad commands
•
Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the transmitter.
Disconnect the security lock jumper!
•
Insert the sensor head into the measurement hole of the
LiCl chamber.
•
Wait at least 20 minutes for the reading to stabilize.
•
Select Cali in the main menu and then RH cal; select Not
changed and then two-point calibration RH 2 point cal.
Change the first point reading with the arrow keys to cor-
respond to the reference humidity and press ENT; pressing
an arrow once changes the reading by 0.05 %RH.