M210300en

Appendix 1: Serial commands

79

ECHO

Turning the serial interface echo ON/OFF

ECHO xxx

xxx

=

ON or OFF

When the echo is off, the commands given through the serial interface or the
prompt > cannot be seen on the display.

When the serial interface is in half-duplex mode, the echo is always off. Even
then the ECHO command can indicate that echo is on.

INTV

Setting the output interval for the RUN mode

INTV xxx yyy

xxx

=

output interval (0...255)
0: no pause between outputs

yyy

=

unit (s, min or h)

Sets the output interval when the transmitter outputs measurement readings to
a peripheral device.

For example, the currently valid settings are output with:

>INTV
Output intrv. : 0 min

When this is changed into 10 minutes, the command is:

>INTV 10
Output intrv. : 10 min

The unit is changed into seconds with:

>INTV S
Output intrv. : 10 s

The change can also be done with one command:

>INTV 10 S
Output intrv. : 10 s