Vaisala HMP240 User Manual
Page 54

HMP240
User's Guide
M210300en
52
6.2.2
Two point temperature calibration
6.2.2.1
With serial commands
•
Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.
Disconnect the security lock jumper!
•
Check the transmitter against the reference.
•
For the humidity sensor head: give command CT
enter the first point value and press
>CT
T : xx.x Ref1 ? yy.y
Press any key when ready
•
If you want to see how the sensor stabilizes to the refer-
ence temperature, enter c
ence:
T : 0.90 Ref1 ? c
T : 0.55 Ref1 ? c
T : 0.55 Ref1 ? 0.0
Press any key when ready...
•
For the additional temperature sensor head: give command
CTA
>CTA
Ta: -20.07 Ref1 ? -20.08
Press any key when ready...
•
Change the temperature and check the transmitter again
against the reference.
•
Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.
•
Press any key, enter the second point value and press
T (or Ta) : xx.x Ref2 ? yy.y
•
The stabilization of the sensor can be monitored by
entering c