HMP240
User's Guide

M210300en

36

5.4.1.2

Selecting and scaling the analogue output quantities

Disconnect the security lock jumper!

ASEL xxx yyy

xxx

=

channel 1's quantity

yyy

=

channel 2's quantity (RH, T, Td, dT, Abs, Mix or Tw)

Example: RH selected on channel 1 and temperature on channel 2

>ASEL RH T
Ch1

(RH)

lo

0.000 %RH ?

Ch1

(RH)

hi

100.000 %RH

?

Ch2

(T )

lo

-40.000 'C

?

Ch2

(T )

hi

+160.000 'C

?

5.4.1.3

Scaling the analogue outputs

Disconnect the security lock jumper!

ASCL

Example: relative humidity is scaled in the range 0...100 %RH and tem-

perature in the range -40...+160 °C

>ASCL
Ch1

(RH)

lo

0.000 %RH ?

Ch1

(RH)

hi

100.000 %RH

?

Ch2

(T )

lo

0.000 'C ? -40

Ch2

(T )

hi

100.000 'C

? 160

5.4.2 Output via the serial bus

5.4.2.1

Starting the measurement output

R

Starts output of measurements to the peripheral devices (RUN mode); the only
command that can be used is S (stop). The output mode can be changed with
command FORM (see Appendix 1).

5.4.2.2

Stopping the measurement output

S

Ends the RUN mode; after this command all other commands are available.