HMP240 series

M210300en

User's Guide

37

5.4.2.3

Outputting the reading once

SEND in STOP mode

or

SEND aa

in POLL mode

aa =

address of the transmitter when more than one transmitter
is connected to a serial bus (0...99)

The output format depends on which parameters the transmitter can output.

Output types:

"Td=999.9 'C",
"RH=999.9 %RH T=999.9 'C Td=9999.9 'C",
"Td=999.9 x=999.9 g/kg",
"RH=999.9 %RH T=999.9 'C Td=9999.9 'C a=9999.9 g/m3
x=9999.9 g/kg Tw=999.9 'C",

The output mode can be changed with command FORM (see Appendix 1).

5.4.2.4

Setting the output interval for the RUN mode

INTV xxx yyy

xxx

=

output interval (0...255)
(0 = no pause between outputs)

yyy

=

unit (s, min or h)

Example: output interval is changed into 10 minutes

>INTV 10 min
Output intrv. : 10 min

5.4.2.5

Serial bus settings

SERI b p d s x

b =

bauds (300, 600, 1200, 2400, 4800, 9600)

p =

parity (n = none, e = even, o = odd)

d =

data bits (7 or 8)

s =

stop bits (1 or 2)

x =

duplex (H = half, F = full)

The settings can be changed one parameter at a time or all parameters at once: